fbpx Sequential – Injection Stripping Analysis(SISA) of Nifuroxime Using DNA-Modified Glassy – Carbon Electrodes |ARAB AMERICAN UNIVERSITY
Contact information for Technical Support and Student Assistance ... Click here

Sequential – Injection Stripping Analysis(SISA) of Nifuroxime Using DNA-Modified Glassy – Carbon Electrodes

Authors: 
Nizam Diab
Ali Z. Abu Zuhri
Wolfgang Schuhmann
Journal Name: 
Bioelectrochemistry
Volume: 
61
Issue: 
1
Pages From: 
57
To: 
63
Date: 
Friday, October 31, 2003
Keywords: 
NifuroximeModified electrodeSequential-injection stripping analysisDNA-modified electrode
Abstract: 
The voltammetric behavior of nifuroxime was investigated comparing stationary voltammetric methods with the recently proposed sequential-injection stripping analysis (SISA), by using cyclic voltammetry (CV) and differential-pulse voltammetry at bare and DNA-modified glassy carbon (GC) electrodes. In cyclic voltammetry, reduction of nifuroxime at DNA-modified electrodes gives rise to a well-defined peak, and in contrast to bare GC surfaces, a re-oxidation peak could be observed. Optimization of the pre-concentration process at the DNA-modified surface led to a significant enhancement of the voltammetric current response, a better defined peak shape and an improved dynamic range. Based on this optimized voltammetric procedure, SISA has been evaluated for the determination of nifuroxime. The flow-system significantly facilitates the regeneration of the DNA-modified electrode surface, hence diminishing problems related to accumulation and memory effects. The linear detection range could be extended to 65 μM with a detection limit (3 s) of 0.68 μM, which corresponds to an absolute amount of 21 ng nifuroxime.